  • 產品名稱:便攜式氧分析儀DF130E

  • 產品型號:DF130E
  • 產品廠商:美國DELETAF
  • 產品文檔:
你添加了1件商品 查看購物車
簡單介紹:
美國DELETAF便攜式氧分析儀DF130,美國AII氧氣分析儀總代理,量程:0-100%GPR-1200/1500/2500/1300/1600, 微量氧氣分析儀EAccuracy: Standard Models: the greater of +/-3% Reading or 0.5% of RangeHigh Resolution Models: the greater of +/-3%
詳情介紹:
便攜式氧分析儀DF130E
Accuracy: (at constant conditions)
Standard Models: the greater of +/-3% Reading or 0.5% of Range
High Resolution Models: the greater of +/-3% of Reading or
0.02% of Range (except for the 0- 100 ppm range: the greate
+/-3% of reading or 50ppb; for the 0-50 ppm range: the
greater of +/-3% of reading or 10ppb)

微量氧氣分析儀DF-130E系列
Oxygen Sensitivity:
Minimum Detectable Change 3 ppb (310E-H0050M model)
Low Detection Limit 3 ppb (310E-H0050M model)
Response Time
Responds instantaneously to O2 change. Typically less than
10 seconds to read 90% of a step change.
(Equilibrium time depends on specific conditions.)
Range: Ranges are available
from 0.5ppm to 25%
Ambient Operating Temperature:
32º to 113ºF (0º to 45ºC)
Background Gas Compatibility
Basic Sensor:
All inert and passive gases including N2, H2, CO, freons,
hydrocarbons, etc.
Sensor with Stab-EL Option:
Neutralizes trace contaminants including acids such as
CO2, H2S, Cl2, NOX, SOX, etc. (Consult Delta F for concentration limits)
Gas Sample Conditions
Sample Pressure
Operating Limits:
0.2 to 1.0 psig (1.03 to 1.08 BarA) - Standard
15-25 psig with welded sample inlet (orifice restricted)
2.0 psi vacuum to 0.2 psig (0.88 to 1.03 BarA) use pump
1.0 to 10 psig (1.08 to 1.7 BarA) use valve (optional) or
regulator (optional)
Above 10 psig (1.7 BarA) use regulator
Sensor overpressure damage limit: 10 psig (1.7 BarA)
Return Pressure: Atmospheric Vent (optimal)
Limits: +5 psig (1.36 BarA) to -5 psig (0.67 BarA)
Flow Rate: 1.0 to 2.0 SCFH (0.5 to 1.0 slpm)
Temperature (Gas Sample): 0º to 150º F (-17.8º to 66ºC)
Moisture: No limits (avoid condensation)
Oil/Solvent Mist: <0.5 mg/ft3 (standard)
>0.5 mg/ft3 - use filter
Solid Particles: <2 mg/ft3 (standard)
>2 mg/ft3 - use filter
 
 
Gas Flow System
Construction Materials: 300 Series stainless steel
Gas Connections: 1/8" compression tube fittings
1/4" VCR compatible (optional
except standard for 310E-H0050M)
Construction:
Enclosure: NEMA 1 - General Purpose
Remote NEMA 4, NEMA 7 or sensor bracket (optional
Weight: 10 lbs. (4.52 kg) (NEMA 1)
Dimensions: 8.32" W x 7.75" H x 7.91" D; (21.1 cm W x 19.7 cm H x 20.1 cm D)
Electrical
Power Input: 22-28 VDC, 1 Amp (max) or 110 VAC or 220
Output Signals: Non-Isolated 0-5, 10 VDC Isolated 4 to 20 mADC (optional
User adjustable to 10% of Full Scale to Full Scale (Std Res)
User adjustable to 1% of Full Scale to Full Scale (High Res)
User selectable Output Freeze during Calibration
Alarms, audible/visual: 4 Oxygen (optional); (adjustable set-point)
Electrolyte Condition (standard)
Temperature (optional)  Low Flow (optional)
Alarm Relays: 4 independently assignable
to Alarms, In-Calibration, Sensor Off and Expanded Range Scale
Alarm Relay Rating: 0.3 Amps at 30 VDC Failsafe Action
Back Lighted Display: Supertwist LCD graphics
Certifications: CE Conformance
Dimensions General Purpose - NEMA 1 Enclosure
 